Calorimeters Series |
 |
Proximate Analyzer |
 |
Elemental Analyzer |
 |
Ash Fusion Tester |
|
|
|
 |
SDC5015 Calorimeter |
|
| |
|
Standard layout:
SDC5015 mainframe, Dedicated water tank, Lenovo computer, Laser printer
Applicable Scope:
It is applicable to measure the calorific value of coal, coke, petroleum, cement black meal and other combustibles.
Quality Certification:
ASTM D 5865-2007a Standard Test Method for Gross Calorific Value of Coal and Coke
GB/T213-2008 Determination of calorific value of coal |
Technical parameter:
Single Complete Testing Duration |
<15min |
Precision |
RSD≤0.1% |
Stability |
Heat Capacity Variation in 3 months≤0.20 % |
Ambient Temperature |
The test results confirm to demands when temperature during(10-30)℃ |
Accuracy |
The mean value and standard value deviation is less than 50J/g after 5 caloric value test of benzoic acid . |
Max Test Range |
42000J /g |
Temperature Resolution |
0.0001 K |
Sample Weight |
0.5-1.5g |
the Water Weight of External Barrel |
40kg |
New type of Oxygen Bomb |
Yes |
Oxygen Bomb 、 Identification |
Automatic |
CAN Bus Interface |
Yes |
Self-diagnostic Function |
Yes |
Complete Appliance Crust Shaped |
No |
Max Power |
1.5kW |
Voltage |
AC220±10 ﹪ 50Hz±1Hz |
Mainframe Size |
440×565×400mm |
Net Weight |
63kg |
|
|